Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Hexachlorocyclopentadiene |
2. | 2,4,5,6-TCMX |
3. | cis-Diallate |
4. | trans-Diallate |
5. | α-BHC |
6. | Hexachlorobenzene |
7. | γ-BHC |
8. | β-BHC |
9. | δ-BHC |
10. | Heptachlor |
11. | Aldrin |
Peaks | |
---|---|
12. | Isodrin |
13. | Heptachlor epoxide |
14. | 2,4'-DDE |
15. | trans-Chlordane |
16. | cis-Chlordane |
17. | Endosulfan I |
18. | 4,4'-DDE |
19. | 2,4'-DDD |
20. | Dieldrin |
21. | Chlorobenzilate |
22. | Endrin |
Peaks | |
---|---|
23. | 2,4'-DDT |
24. | 4,4'-DDD |
25. | Endosulfan II |
26. | Kepone |
27. | Endrin aldehyde |
28. | 4,4'-DDT |
29. | Endosulfan sulfate |
30. | Methoxychlor |
31. | Endrin ketone |
32. | Mirex |
33. | DCB |
Columns | Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.50 µm (cat.# 11325) |
---|---|
Standard/Sample | Organochlorine pesticide mix AB #2 (cat.# 32292) |
Chlorobenzilate (cat.# 32211) | |
Diallate (cis & trans) (cat.# custom) | |
Hexachlorobenzene (cat.# 32231) | |
Hexachlorocyclopentadiene (cat.# 32232) | |
Isodrin (cat.# custom) | |
Kepone (cat.# custom) | |
Mirex (cat.# custom) | |
2,4'-DDD (cat.# 32098) | |
2,4'-DDE (cat.# 32099) | |
2,4'-DDT (cat.# 32200) | |
TCMX (cat.# 32027) | |
Decachlorobiphenyl (BZ #209) (cat.# 32029) | |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.75 min) |
Inj. Temp.: | 275 °C |
Oven | |
Oven Temp.: | 140 °C (hold 1 min) to 240 °C at 30 °C/min (hold 2 min) to 330 °C at 30 °C/min (hold 4 min) |
Carrier Gas | H2, constant pressure |
Linear Velocity: | 51 cm/sec @ 140 °C |
Detector | ECD @ 330 °C |
---|---|
Instrument | Shimadzu 2010 GC |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.