Peaks | |
---|---|
1. | 1,4-Dioxane |
2. | N-Nitrosodimethylamine |
3. | Pyridine |
4. | 2-Fluorophenol (SS) |
5. | Phenol-d6- (SS) |
6. | Phenol |
7. | Aniline |
8. | Bis(2-chloroethyl) ether |
9. | 2-Chlorophenol |
10. | 1,3-Dichlorobenzene |
11. | 1,4-dichlorobenzene-d4 (IS) |
12. | 1,4-Dichlorobenzene |
13. | Benzyl alcohol |
14. | 1,2-Dichlorobenzene |
15. | 2-Methylphenol |
16. | Bis(2-chloroisopropyl) ether |
17. | 4-methylphenol/3-methylphenol |
18. | N-Nitrosodi-N-propylamine |
19. | Hexachloroethylene |
20. | nitrobenzene-d5 (SS) |
21. | Nitrobenzene |
22. | Isophorone |
23. | 2-Nitrophenol |
24. | 2,4-Dimethylphenol |
25. | Bis(2-chloroethoxy)methane |
26. | Benzoic acid |
27. | 2,4-Dichlorophenol |
28. | 1,2,4-Trichlorobenzene |
29. | Naphthalene-d8 (IS) |
30. | Naphthalene |
31. | 4-Chloroaniline |
Column | Rxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602) |
---|---|
Standard/Sample | 8270 MegaMix (cat.# 31850) |
benzoic acid (cat.# 31879) | |
8270 Benzidines Mix (cat.# 31852) | |
Acid Surrogate Mix (4/89 SOW) (cat.# 31025) | |
Revised B/N Surrogate Mix (cat.# 31887) | |
1,4-dioxane (cat.# 31853) | |
SV Internal Standard Mix (cat.# 31206) | |
Conc.: | 10 ng/mL on-column concentration |
Injection | |
Inj. Vol.: | 1.0 µL pulsed splitless (hold 0.15 min) |
Liner: | Drilled Uniliner (hole near bottom) (cat.# 20756) |
Inj. Temp.: | 250 °C |
Pulse Pressure: | 30 psi (206.8kPa) |
Pulse Time: | 0.2 min |
Purge Flow: | 60 mL/min |
Oven | |
Oven Temp.: | 50 °C (hold 0.5 min) to 260 °C at 20 °C/min to 280 °C at 5 °C/min to 330 °C at 20 °C/min (hold 1.0 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |