Peaks | |
---|---|
1. | BDE-10 |
2. | BDE-7 |
3. | BDE-8 |
4. | BDE-11 |
5. | BDE-12 |
6. | BDE-13 |
7. | BDE-15 |
8. | BDE-30 |
9. | BDE-32 |
10. | BDE-17 |
11. | BDE-25 |
12. | BDE-28 |
13. | BDE-33 |
Peaks | |
---|---|
14. | BDE-35 |
15. | BDE-37 |
16. | BDE-75 |
17. | BDE-49 |
18. | BDE-71 |
19. | BDE-47 |
20. | BDE-66 |
21. | BDE-77 |
22. | BDE-100 |
23. | BDE-119 |
24. | BDE-99 |
25. | BDE-116 |
26. | BDE-118 |
Peaks | |
---|---|
27. | BDE-85 |
28. | BDE-155 |
29. | BDE-126 |
30. | BDE-154 |
31. | BDE-153 |
32. | BDE-138 |
33. | BDE-166 |
34. | BDE-183 |
35. | BDE-181 |
36. | BDE-190 |
37. | BDE-208 |
38. | BDE-207 |
39. | BDE-206 |
40. | BDE-209 |
Column | Rtx-1614, 30 m, 0.25 mm ID, 0.10 µm (cat.# 10295) |
---|---|
Standard/Sample | 100-300 ppb PBDE PAR solution (cat.# EO-5113, Cambridge Isotope Laboratories Inc.) |
500 ppb decabromodiphenyl ether (cat.# BDE-209, Wellington Laboratories) | |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Cyclo double taper (4 mm) (cat.# 20896) |
Inj. Temp.: | 300 °C |
Oven | |
Oven Temp.: | 100 °C (hold 3 min) to 320 °C at 5 °C/min (hold 15 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 20 cm/sec @ 100 °C |
Detector | µ-ECD @ 340 °C |
---|