Peaks | |
---|---|
1. | Naphthalene |
2. | 2-Methylnaphthalene |
3. | 1-Methylnaphthalene |
4. | Acenaphthylene |
5. | Acenaphthene |
6. | Fluorene |
7. | Phenanthrene |
8. | Anthracene |
9. | Fluoranthene |
10. | Pyrene |
11. | Benz[a]anthracene |
12. | Chrysene |
13. | Triphenylene |
Column | Rxi-17Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123) |
---|---|
Standard/Sample | PAH Supplement Mix for Method 8100 (cat.# 31857) |
EPA Method 8310 PAH mixture (cat.# 31841) | |
Triphenylene (custom) | |
Diluent: | Dichloromethane |
Conc.: | 10 ppm |
Injection | |
Inj. Vol.: | 0.5 µL splitless (hold 1.75 min) |
Liner: | Auto SYS XL PSS split/splitless w/wool (cat.# 21718) |
Inj. Temp.: | 320 °C |
Purge Flow: | 75 mL/min |
Oven | |
Oven Temp.: | 65 °C (hold 0.5 min) to 220 °C at 15 °C/min to 330 °C at 4 °C/min (hold 15 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.0 mL/min |
Detector | FID @ 320 °C |
---|---|
Instrument | PE Clarus 600 GC |
Acknowledgement | Instrument provided by PerkinElmer |