Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
---|---|
Standard/Sample | 8270 MegaMix (cat.# 31850) |
8270 Benzidines mix (cat.# 31852) | |
Benzoic acid (cat.# 31879) | |
Revised B/N surrogate mix (cat.# 31888) | |
Acid surrogate mix (4/89 SOW) (cat.# 31063) | |
Revised SV internal standard mix (cat.# 31886) | |
Appendix IX mix #1 (cat.# 31625) | |
Appendix IX mix #2 (cat.# 31806) | |
Diluent: | Dichloromethane |
Conc.: | 0.5 µg/mL (IS/SS 20 μg/mL) |
Injection | |
Inj. Vol.: | 1 µL pulsed splitless (hold 0.59 min) |
Liner: | Premium 4 mm single taper/gooseneck w/wool (cat.# 23303.1) |
Inj. Temp.: | 270 °C |
Pulse Pressure: | 30 psi (206.8kPa) |
Pulse Time: | 0.64 min |
Purge Flow: | 100 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 280 °C at 25 °C/min to 320 °C at 5 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 276 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 70 eV |
Solvent Delay Time: | 2.19 min |
Tune Type: | DFTPP |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |
Scan Rate: | 5.36 scans/sec |
Instrument | Agilent 7890A GC & 5975C MSD |
Notes | 7890 Siltek-treated EZ Twist Top split/splitless injection port (cat.# 22178) Flip Seal dual Vespel ring inlet seal (cat.# 23411) |