Peaks | |
---|---|
1. | N-Nitrosodimethylamine-d6 |
2. | N-Nitrosodimethylamine |
3. | N-Nitrosomethylethylamine |
4. | N-Nitrosodiethylamine |
5. | N-Nitrosodipropylamine-d14 |
6. | N-Nitrosodipropylamine |
7. | N-Nitrosopyrrolidine |
8. | N-Nitrosopiperidine |
9. | N-Nitrosodibutylamine |
Column | Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868) |
---|---|
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029) | |
with universal Press-Tight connectors (cat.# 20429) | |
Standard/Sample | N-Nitrosodimethylamine-d6 (cat.# 33910) |
N-Nitrosodi-n-propylamine-d14 (cat.# 33911) | |
Nitrosamine calibration mix, method 521 (cat.# 31898) | |
Diluent: | Dichloromethane |
Conc.: | 10 ng/mL |
Injection | |
Inj. Vol.: | 10 µL splitless (hold 1 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.5) |
Inj. Temp.: | 250 °C |
Purge Flow: | 80 mL/min |
Oven | |
Oven Temp.: | 38 °C (hold 1 min) to 160 °C at 12 °C/min to 200 °C at 5 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | Agilent 5975C | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
SIM Program: | |||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
Transfer Line Temp.: | 280 °C | ||||||||||||||||||||||||||||
Analyzer Type: | Quadrupole | ||||||||||||||||||||||||||||
Source Temp.: | 230 °C | ||||||||||||||||||||||||||||
Quad Temp.: | 150 °C | ||||||||||||||||||||||||||||
Solvent Delay Time: | 6.9 min | ||||||||||||||||||||||||||||
Tune Type: | BFB | ||||||||||||||||||||||||||||
Ionization Mode: | EI | ||||||||||||||||||||||||||||
Instrument | Agilent 7890A GC & 5975C MSD |