Peaks | Conc. (ng/mL) | |
---|---|---|
1. | Tetrahydrofuran-d8 (IS) | 100 |
2. | Tetrahydrofuran | 50 |
3. | 1,4-Dioxane-d8 (SS) | 200 |
4. | 1,4-Dioxane | 50 |
5. | N-Nitrosodimethylamine-d6 (SS) | 20 |
6. | N-Nitrosodimethylamine | 2.5 |
7. | N-Nitrosomethylethylamine | 5.0 |
Peaks | Conc. (ng/mL) | |
---|---|---|
8. | N-Nitrosodiethylamine | 5.0 |
9. | N-Nitrosodi-n-propylamine-d14 (IS) | 10 |
10. | N-Nitrosopyrrolidine | 5.0 |
11. | N-Nitrosodi-n-propylamine | 5.0 |
12. | N-Nitrosomorpholine | 2.5 |
13. | N-Nitrosopiperidine | 5.0 |
14. | N-Nitrosodi-n-butylamine | 5.0 |
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 13653) |
---|---|
using Rxi guard column 10 m, 0.53 mm ID (cat.# 10073) | |
with SGE µ-union | |
Standard/Sample | N-Nitrosodimethylamine-d6 (cat.# 33910) |
1,4-Dioxane-d8 (cat.# 30614) | |
N-Nitrosodi-n-propylamine-d14 (cat.# 33911) | |
Tetrahydrofuran-d8 (cat.# 30112) | |
Nitrosamine calibration mix, Method 521 (cat.# 31898) | |
Appendix IX mix #1, revised (cat.# 32459) | |
Diluent: | Dichloromethane |
Injection | |
Inj. Vol.: | 50 µL splitless (hold 1.5 min) |
Liner: | Custom Premium single taper 4 mm ID liner with 15 mg quartz wool |
Inj. Temp.: | 275 °C |
Purge Flow: | 100 mL/min |
Oven | |
Oven Temp.: | 35 °C (hold 1.5 min) to 50 °C at 50 °C/min (hold 7.1 min) to 320 °C at 11.12 °C/min (hold 1.5 min) |
Carrier Gas | He, flow program |
Flow Program: | 5.08 mL/min (hold 8.9 min) to 2.0 mL/min at 1.0 mL/min/min |
Detector | MS | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Mode: | SIM | ||||||||||||||||||||||||||||
SIM Program: | |||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
Transfer Line Temp.: | 320 °C | ||||||||||||||||||||||||||||
Analyzer Type: | Quadrupole | ||||||||||||||||||||||||||||
Source Temp.: | 230 °C | ||||||||||||||||||||||||||||
Quad Temp.: | 150 °C | ||||||||||||||||||||||||||||
Solvent Delay Time: | 9.9 min | ||||||||||||||||||||||||||||
Tune Type: | BFB | ||||||||||||||||||||||||||||
Ionization Mode: | EI | ||||||||||||||||||||||||||||
Instrument | Agilent 7890A GC & 5975C MSD |