Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

50 µL CSR-LVSI with 5 ng Nitrosamines, THF, and 1,4-Dioxane On-Column

ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 13653)
using Rxi guard column 10 m, 0.53 mm ID (cat.# 10073)
with BGB P/N: 2553LD
Standard/Sample1,4-Dioxane (cat.# 30287)
Nitrosamine calibration mix, Method 521 (cat.# 31898)
Tetrahydrofuran (THF) (cat.# 30414)
Diluent:Dichloromethane
Conc.:0.1 µg/mL
Injection
Inj. Vol.:50 µL splitless (hold 1.5 min)
Liner:Custom Premium single taper with 15 mg quartz wool
Inj. Temp.:275 °C
Purge Flow:100 mL/min
Oven
Oven Temp.:35 °C (hold 1.5 min) to 50 °C at 50 °C/min (hold 7.1 min) to 320 °C at 11.12 °C/min (hold 1.5 min)
Carrier GasHe, constant flow
Flow Rate:5.08 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
19.7042, 46, 71, 72, 78, 80 40
210.2558, 62, 64, 88, 96 40
310.7542, 43, 46, 48, 74, 80 40
411.5042, 44, 56, 88, 102 40
514.0056, 68, 70, 78, 86, 100, 130, 144 25
616.2055, 56, 84, 114 40
717.5099, 116, 141, 158 40
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:9.7 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD