Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Column | Rxi-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950) |
---|---|
Standard/Sample | Tarball 6 - Florida Gulf beach |
Diluent: | Carbon disulfide |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 17.5 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.1 min) to 400 °C at 20 °C/min (hold 1.9 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.75 mL/min |
Detector | FID @ 420 °C |
---|---|
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Actual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc). |
Acknowledgement | Susan Forsyth - Walton County, FL - Tarball samples |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.