Peaks | tR (min) | |
---|---|---|
1. | Isophorone | 6.13 |
2. | 1,3-Dimethyl-2-nitrobenzene (SS) | 7.03 |
3. | Dimethylphthalate | 10.58 |
4. | Acenaphthylene | 10.74 |
5. | Acenaphthene-D10 (IS) | 11.08 |
6. | Diethylphthalate | 12.28 |
7. | Fluorene | 12.39 |
8. | Pentachlorophenol-C13 (IS) | 14.19 |
9. | Phenanthrene-D10 (IS) | 14.60 |
10. | Phenanthrene | 14.66 |
11. | Anthracene | 14.78 |
12. | Dibutyl phthalate | 16.40 |
13. | Pyrene | 18.46 |
Peaks | tR (min) | |
---|---|---|
14. | Butylbenzylphthalate | 20.82 |
15. | Di(2-ethylhexyl)adipate | 21.29 |
16. | Triphenyl phosphate (SS) | 21.38 |
17. | Benzo[a]anthracene | 22.19 |
18. | Chrysene-D12 (IS) | 22.22 |
19. | Chrysene | 22.30 |
20. | Di(2-ethylhexyl)phthalate | 22.84 |
21. | Benzo[b]fluoranthene | 25.45 |
22. | Benzo[k]fluoranthene | 25.53 |
23. | Benzo[a]pyrene-D12 (SS) | 26.27 |
24. | Benzo[a]pyrene | 26.34 |
25. | Indeno[1,2,3-cd]pyrene | 29.25 |
26. | Dibenzo[a,h]anthracene | 29.34 |
27. | Benzo[ghi]perylene | 29.84 |
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
---|---|
Standard/Sample | EPA Method 525.3 PAH cal standard (cat.# 32545) |
EPA Method 525.3 PAH IS mix (cat.# 32547) | |
EPA Method 525.3 PCP IS (cat.# 32548) | |
EPA Method 525.3 surrogate standard (cat.# 32549) | |
Diluent: | Ethyl acetate |
Conc.: | 2 µg/mL |
Injection | |
Inj. Vol.: | 1 µL pulsed splitless (hold 1 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.1) |
Inj. Temp.: | 275 °C |
Pulse Pressure: | 30 psi (206.8kPa) |
Pulse Time: | 1 min |
Purge Flow: | 80 mL/min |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 200 °C at 10 °C/min to 320 °C at 7 °C/min (hold 3 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Extractor | ||||||||
Extractor Lens: | 6 mm ID | ||||||||
Source Temp.: | 350 °C | ||||||||
Quad Temp.: | 200 °C | ||||||||
Solvent Delay Time: | 5 min | ||||||||
Tune Type: | DFTPP | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890B GC & 5977A MSD |