Peaks | tR (min) | Conc. (µg/mL) | |
---|---|---|---|
1. | Naphthalene-d8 (IS) | 3.44 | 2.0 |
2. | Naphthalene | 3.45 | 1.0 |
3. | 2-Methylnaphthalene | 3.86 | 1.0 |
4. | 1-Methylnaphthalene | 3.92 | 1.0 |
5. | 2-Fluorobiphenyl (SS) | 4.08 | 2.0 |
6. | Acenaphthylene | 4.41 | 1.0 |
7. | Acenaphthene-d10 (IS) | 4.49 | 2.0 |
8. | Acenaphthene | 4.51 | 1.0 |
9. | Fluorene | 4.83 | 1.0 |
10. | 2,4,6-Tribromophenol (SS) | 4.97 | 2.0 |
11. | Phenanthrene-d10 (IS) | 5.40 | 2.0 |
12. | Phenanthrene | 5.41 | 1.0 |
13. | Anthracene | 5.45 | 1.0 |
Peaks | tR (min) | Conc. (µg/mL) | |
---|---|---|---|
14. | Fluoranthene | 6.15 | 1.0 |
15. | Pyrene-d10 (SS) | 6.28 | 2.0 |
16. | Pyrene | 6.29 | 1.0 |
17. | p-Terphenyl-D14 (SS) | 6.37 | 2.0 |
18. | Benz[a]anthracene | 7.12 | 1.0 |
19. | Chrysene-d12 (IS) | 7.13 | 2.0 |
20. | Chrysene | 7.15 | 1.0 |
21. | Benzo[b]fluoranthene | 8.17 | 1.0 |
22. | Benzo[k]fluoranthene | 8.21 | 1.0 |
23. | Benzo[a]pyrene | 8.56 | 1.0 |
24. | Perylene-d12 (IS) | 8.64 | 2.0 |
25. | Indeno[1,2,3-cd]pyrene | 10.16 | 1.0 |
26. | Dibenz[a,h]anthracene | 10.20 | 1.0 |
27. | Benzo[ghi]perylene | 10.58 | 1.0 |
Column | Rxi-5Sil MS, 20 m, 0.15 mm ID, 0.15 µm (cat.# 43816) |
---|---|
Standard/Sample | EPA Method 8310 PAH mixture (cat.# 31874) |
Revised SV internal standard mix (cat.# 31886) | |
Revised B/N surrogate mix (cat.# 31888) | |
Acid surrogate mix (4/89 SOW) (cat.# 31063) | |
Diluent: | Dichloromethane |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 20:1) |
Liner: | Topaz 4 mm single taper w/wool (cat.# 23303) |
Inj. Temp.: | 275 °C |
Oven | |
Oven Temp.: | 60 °C (hold 0.7 min) to 285 °C at 39.8 °C/min to 305 °C at 4.3 °C/min to 320 °C at 28.5 °C/min (hold 3.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS | ||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Mode: | SIM | ||||||||||||||||||||||||||||||||||||||||||||
SIM Program: | |||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||
Transfer Line Temp.: | 280 °C | ||||||||||||||||||||||||||||||||||||||||||||
Analyzer Type: | Quadrupole | ||||||||||||||||||||||||||||||||||||||||||||
Source Type: | Extractor | ||||||||||||||||||||||||||||||||||||||||||||
Extractor Lens: | 9 mm ID | ||||||||||||||||||||||||||||||||||||||||||||
Source Temp.: | 330 °C | ||||||||||||||||||||||||||||||||||||||||||||
Quad Temp.: | 180 °C | ||||||||||||||||||||||||||||||||||||||||||||
Solvent Delay Time: | 1 min | ||||||||||||||||||||||||||||||||||||||||||||
Tune Type: | DFTPP | ||||||||||||||||||||||||||||||||||||||||||||
Ionization Mode: | EI | ||||||||||||||||||||||||||||||||||||||||||||
Instrument | Agilent 7890B GC & 5977A MSD | ||||||||||||||||||||||||||||||||||||||||||||
Notes | Fast SIM analysis of 16 priority PAHs plus the methylnaphthalenes in a 120 V oven equipped with the GC Accelerator kit (cat.# 23849) (injected 1 µg/mL = 0.05 ng on-column). |