Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Resprep CarboPrep Plus SPE Cartridge (Lot 1221 Comparison)

ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
Standard/Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Diluent:Hexane
Conc.:2 ppb
Injection
Inj. Vol.:4 µL splitless (hold 0.75 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:90 °C (hold 0.5 min) to 330 °C at 25 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
DetectorMicro-ECD @ 340 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
Sample PreparationConditioned a Resprep CarboPrep Plus SPE cartridge (cat.# 25845) by adding one cartridge volume of hexane:acetone (90:10) and letting it stand for five minutes before drawing the solvent down to frit level. Then, 1 mL of sample extract was loaded onto the cartridge and eluted with 9 mL of hexane:acetone (90:10).