Column | Rxi-SVOCms, 30 m, 0.25 mm ID, 0.50 µm (cat.# 16638) |
---|---|
Standard/Sample | 8270/Appendix IX kit (cat.# 31815) |
Revised B/N surrogate mix (cat.# 31888) | |
Acid surrogate mix (cat.# 31063) | |
Diluent: | Dichloromethane |
Conc.: | 20 µg/mL (2 ng on-column) |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23303) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 280 °C at 12.4 °C/min to 330 °C at 3.3 °C/min (hold 3.65 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Drawout Plate: | 6 mm ID | ||||||||
Source Temp.: | 330 °C | ||||||||
Quad Temp.: | 180 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 2 min | ||||||||
Tune Type: | DFTPP | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Sample Preparation | Samples were aliquoted into amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842). |