Semivolatiles SVOC Additions Standard on Rxi-SVOCms

GC_EV1506
PeakstR (min)Conc.
(µg/mL)
Quantifier IonQualifier Ion
1.Acrylamide3.7672007155
2.Decane5.4432005771
3.Indene6.07200115116
4.O,O,O-Triethyl phosphorothioate7.08200198121
5.a-Terpineol7.5620093136
6.Quinoline8.071200129102
7.2,3-Dichloroaniline9.05200161163
8.Hexadecane10.912007185
9.Zinophos11.0120014397
10.Sulfotep11.49200322.2202
11.Phorate11.7320075121
PeakstR (min)Conc.
(µg/mL)
Quantifier IonQualifier Ion
12.Dimethoate11.9820087125
13.Octadecane12.352007185
14.Disulfoton12.532008897
15.Dinoseb12.54200211163
16.Methyl parathion13.13200263109
17.Ethyl parathion13.7320029197
18.Benzidine14.65200184185
19.Famphur15.74200218125
20.3,3'-Dichlorobenzidine16.92200252254
21.4,4'-Methylenebis(2-chloroaniline)16.92200266231
22.6-Methylchrysene18.21200242.2241
23.Dibenz(a,h)acridine24.62200279.1139
ColumnRxi-SVOCms, 30 m, 0.25 mm ID, 0.50 µm (cat.# 16638)
Standard/SampleSVOC Additions standard  (cat.# 31909)
Diluent:Dichloromethane
Conc.:200 µg/mL (2 ng on-column)
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23303)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 1 min) to 285 °C at 16 °C/min to 305 °C at 1.65 °C/min to 330 °C at 8.8 °C/min (hold 4.55 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.050-5003.6
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:6 mm ID
Source Temp.:330 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Solvent Delay Time:2 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Sample PreparationStandard was diluted to a concentration of 200 ppm with dichloromethane in amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842).
Notes* Denotes p-tert-butylcatechol (preservative)
AcknowledgementChris Rattray