Peaks | tR (min) | |
---|---|---|
1. | (IS) 1,4-Dioxane-d8 | 1.87 |
2. | N-Nitrosodimethylamine | 2.00 |
3. | Pyridine | 2.03 |
4. | (SS) 2-Fluorophenol | 2.67 |
5. | (SS) Phenol-d6 | 3.29 |
6. | Phenol | 3.30 |
7. | Aniline | 3.36 |
8. | Bis(2-chloroethyl) ether | 3.40 |
9. | 2-Chlorophenol | 3.46 |
10. | 1,3-Dichlorobenzene | 3.59 |
11. | (IS) 1,4-Dichlorobenzene-D4 | 3.63 |
12. | 1,4-Dichlorobenzene | 3.65 |
13. | Benzyl alcohol | 3.72 |
14. | 1,2-Dichlorobenzene | 3.78 |
15. | 2-Methylphenol | 3.80 |
16. | Bis(2-Chloroisopropyl)ether | 3.84 |
17. | 4-Methylphenol | 3.93 |
18. | 3-Methylphenol | 3.93 |
19. | N-Nitrosodi-N-propylamine | 3.95 |
20. | Hexachloroethane | 4.07 |
21. | (SS) Nitrobenzene-D5 | 4.10 |
22. | Nitrobenzene | 4.11 |
23. | Isophorone | 4.32 |
24. | 2-Nitrophenol | 4.40 |
25. | 2,4-Dimethylphenol | 4.42 |
26. | Benzoic acid | 4.46 |
27. | Bis(2-chloroethoxy)methane | 4.51 |
28. | 2,4-Dichlorophenol | 4.61 |
29. | 1,2,4-Trichlorobenzene | 4.70 |
30. | (IS) Naphthalene-D8 | 4.76 |
31. | Naphthalene | 4.78 |
Peaks | tR (min) | |
---|---|---|
32. | 4-Chloroaniline | 4.82 |
33. | Hexachlorobutadiene | 4.89 |
34. | 4-Chloro-3-methylphenol | 5.26 |
35. | 2-Methylnaphthalene | 5.43 |
36. | 1-Methylnaphthalene | 5.53 |
37. | Hexachlorocyclopentadiene | 5.59 |
38. | 2,4,6-Trichlorophenol | 5.70 |
39. | 2,4,5-Trichlorophenol | 5.73 |
40. | (SS) 2-Fluorobiphenyl | 5.79 |
41. | 2-Chloronaphthalene | 5.91 |
42. | 2-Nitroaniline | 6.00 |
43. | 1,4-Dinitrobenzene | 6.13 |
44. | Dimethyl phthalate | 6.18 |
45. | 1,3-Dinitrobenzene | 6.20 |
46. | 2,6-Dinitrotoluene | 6.24 |
47. | 1,2-Dinitrobenzene | 6.29 |
48. | Acenaphthylene | 6.31 |
49. | 3-Nitroaniline | 6.40 |
50. | (IS) Acenaphthene-D10 | 6.45 |
51. | Acenaphthene | 6.48 |
52. | 2,4-Dinitrophenol | 6.50 |
53. | 4-Nitrophenol | 6.55 |
54. | 2,4-Dinitrotoluene | 6.63 |
55. | Dibenzofuran | 6.65 |
56. | 2,3,5,6-Tetrachlorophenol | 6.73 |
57. | 2,3,4,6-Tetrachlorophenol | 6.77 |
58. | Diethyl phthalate | 6.88 |
59. | 4-Chlorophenyl phenyl ether | 6.99 |
60. | Fluorene | 6.99 |
61. | 4-Nitroaniline | 7.00 |
62. | 4,6-Dinitro-2-methylphenol | 7.03 |
Peaks | tR (min) | |
---|---|---|
63. | N-Nitrosodiphenylamine | 7.10 |
64. | N,N-Diphenylhydrazine | 7.15 |
65. | (SS) 2,4,6-Tribromophenol | 7.23 |
66. | 4-Bromophenyl phenyl ether | 7.47 |
67. | Hexachlorobenzene | 7.53 |
68. | Pentachlorophenol | 7.72 |
69. | (IS) Phenanthrene-D10 | 7.92 |
70. | Phenanthrene | 7.94 |
71. | Anthracene | 7.99 |
72. | Carbazole | 8.15 |
73. | di-n-Butyl phthalate | 8.49 |
74. | Fluoranthene | 9.12 |
75. | Benzidine | 9.24 |
76. | (SS) Pyrene-D10 | 9.32 |
77. | Pyrene | 9.34 |
78. | (SS) p-Terphenyl-d14 | 9.49 |
79. | 3,3'-Dimethylbenzidine | 9.98 |
80. | Butyl benzyl phthalate | 10.00 |
81. | Bis(2-ethylhexyl) adipate | 10.09 |
82. | 3,3'-Dichlorobenzidine | 10.62 |
83. | Benz[a]anthracene | 10.66 |
84. | (IS) Chrysene-D12 | 10.67 |
85. | Chrysene | 10.71 |
86. | Bis(2-ethylhexyl) phthalate | 10.71 |
87. | Di-n-octyl phthalate | 11.68 |
88. | Benzo[b]fluoranthene | 12.30 |
89. | Benzo[k]fluoranthene | 12.34 |
90. | Benzo[a]pyrene | 12.89 |
91. | (IS) Perylene-D12 | 13.00 |
92. | Indeno[1,2,3-cd]pyrene | 15.32 |
93. | Dibenz[a,h]anthracene | 15.40 |
94. | Benzo[ghi]perylene | 15.95 |
Column | Rxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623) |
---|---|
Standard/Sample | 8270 MegaMix standard (cat.# 31850) |
8270 Benzidines mix (cat.# 31852) | |
Benzoic acid (cat.# 31879) | |
Revised SV internal standard mix (cat.# 31886) | |
Revised B/N surrogate mix (cat.# 31888) | |
Acid surrogate mix (cat.# 31063) | |
Diluent: | Dichloromethane |
Conc.: | 20 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Topaz 4.0 mm ID single taper inlet liner with wool (cat.# 23303) |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 12 mL/min |
Oven | |
Oven Temp.: | 60 °C (hold 0.5 min) to 285 °C at 25 °C/min to 305 °C at 3 °C/min to 330 °C at 20 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Drawout Plate: | 6 mm ID | ||||||||
Source Temp.: | 330 °C | ||||||||
Quad Temp.: | 180 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Tune Type: | DFTPP | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Sample Preparation | Samples were aliquoted into amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842). |