Peaks | |
---|---|
1. | Pyridine |
2. | Pentachlorophenol |
Column | Rxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623) |
---|---|
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Topaz 4.0 mm ID single taper inlet liner with wool (cat.# 23303) |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 12 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.5 min) to 280 °C at 20 °C/min to 330 °C at 6 °C/min (hold 4 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Type: | Extractor |
Extractor Lens: | 6 mm ID |
Source Temp.: | 330 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 70 eV |
Solvent Delay Time: | 1.55 min |
Tune Type: | DFTPP |
Ionization Mode: | EI |
Instrument | Agilent 7890A GC & 5975C MSD |
Sample Preparation | Samples were aliquoted into amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842). |