Priority PAH Pollutants on Rxi-SVOCms (SIM)

GC_EV1609
PeakstR (min)
1.Naphthalene6.27
2.2-Methylnaphthalene7.09
3.1-Methylnaphthalene7.20
4.Biphenyl7.65
5.2,6-Dimethylnaphthalene7.84
6.Acenaphthylene8.17
7.Acenaphthene8.38
8.2,3,5-Trimethylnaphthalene8.85
9.Fluorene9.01
10.Dibenzothiophene10.02
11.Phenanthrene10.18
PeakstR (min)
12.Anthracene10.24
13.1-Methylphenanthrene10.94
14.Fluoranthene11.64
15.Pyrene11.91
16.Benz[a]anthracene13.45
17.Chrysene13.50
18.Benzo[b]fluoranthene15.13
19.Benzo[k]fluoranthene15.18
20.Benzo[a]pyrene15.69
21.Indeno[1,2,3-cd]pyrene17.77
22.Dibenz[a,h]anthracene17.82
23.Benzo[ghi]perylene18.26
The internal standard and surrogate standard mass on column is 20 pg.
ColumnRxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623)
Standard/SampleCustom PAH SIM standard
Diluent:Dichloromethane
Conc.:40 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID single taper inlet liner with wool (cat.# 23303)
Inj. Temp.:250 °C
Split Vent Flow Rate:24 mL/min
Oven
Oven Temp.:40 °C (hold 0.5 min) to 280 °C at 20 °C/min to 330 °C at 6 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
15.00127.05, 128.05, 129.00 25
26.75115.10, 139.00, 141.00, 142.05 25
37.47141.00, 152.00, 153.05, 154.05, 155.05, 156.10, 162.10, 164.10 15
48.03150.00, 151.05, 152.05, 153.05, 154.10, 162.10, 164.10 15
58.66153.05, 155.10, 163.05, 164.10, 165.05, 166.05, 169.10, 170.10 15
69.62139.00, 151.95, 176.10, 177.10, 178.10, 179.10, 183.95, 185.00, 188.10, 189.10 15
710.71189.05, 190.05, 191.10, 192.10 25
811.37200.10, 201.10, 202.10, 203.05 20
911.81200.10, 201.05, 202.05, 203.05 20
1012.84114.00, 120.00, 226.10, 227.10, 228.10, 229.10, 240.10 15
1114.48126.00, 132.00, 250.10, 252.10, 253.10, 264.00 15
1215.51126.00, 132.00, 250.05, 252.05, 253.05, 264.00 20
1316.95134.50, 137.95, 139.00, 274.05, 276.10, 277.10, 278.10, 279.10 20
1418.10138.00, 139.00, 274.05, 276.10, 277.10, 278.1 25
1518.80150.00, 151.00, 156.00, 600.00, 302.00, 312.00 25
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:330 °C
Quad Temp.:150 °C
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Sample Preparation200 ppm standard diluted 5x, then analyzed at 20:1 split. Samples were aliquoted into amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842).