Peaks | |
---|---|
1. | Acetaldehyde |
2. | Methanol |
3. | Acetone |
4. | Ethyl formate |
5. | Ethanol |
6. | Ethyl Acetate |
Peaks | |
---|---|
7. | n-Propanol |
8. | sec-Butanol |
9. | Isobutanol |
10. | Active amyl alcohol |
11. | Isoamyl alcohol |
12. | n-Amyl alcohol |
Column | 5% Carbowax® 20M, 80/120 mesh on CarboBlack B, SilcoSmooth™ Tubing, 2 m, 1/8 in. OD, 2 mm ID (cat.# 80105-800) |
---|---|
Standard/Sample | |
Conc.: | neat |
Injection | |
Inj. Vol.: | 0.5 µL packed on-column |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 65 °C (hold 5 min) to 150 °C at 4 °C/min |
Carrier Gas | N2, constant flow |
Flow Rate: | 20 mL/min |
Detector | FID @ 250 °C |
---|---|
Notes | FID sensitivity: 1.28 x 10-10 AFS |