Peaks | |
---|---|
1. | (possible 3-methyl-2(5H)-furanone) |
2. | valeric acid + susquiterpene |
3. | Linalool oxide |
4. | Susquiterpene |
5. | Isobutyl decanoate |
6. | 1-Decanol |
7. | β-Citronellol |
8. | Susquiterpene |
9. | methyl salicylate + unknown |
10. | Siloxane |
11. | Methyl dodecanoate |
12. | hexyl octanoate + |
13. | 2-Tridecanone |
14. | trans-2-trans-4-Decadienol |
15. | 2-Phenylethyl acetate |
Peaks | |
---|---|
16. | Hexanoic acid |
17. | Ethyl dodecanoate |
18. | Unknown |
19. | Isoamyl decanoate |
20. | Benzyl Alcohol |
21. | Unknown |
22. | 2-Phenylethanol |
23. | 2-Ethylhexanoic acid |
24. | Heptanoic acid |
25. | Dodecanol |
26. | Phenol |
27. | γ-Nonalactone |
28. | Octanoic Acid |
29. | Siloxane |
30. | Ethyl cinnamate |
Peaks | |
---|---|
31. | γ-Decalactone + unknown |
32. | Nonanoic acid |
33. | Phenylethyl hexanoate |
34. | Diethyl nonanedioate |
35. | Methyl hexadecanoate |
36. | Dimethylnaphthalene |
37. | Ethyl hexadecanoate |
38. | Decanoic acid |
39. | Ethyl 9-hexadecenoate |
40. | Geranic acid |
41. | Phenylethyl octanoate |
42. | Dodecanoic acid |
43. | Ethyl linoleate |
44. | Diisobutyl phthalate |
45. | Ethyl linolenate |
46. | Phenylethyl decanoate |
Column | Stabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752) |
---|---|
Injection | |
Inj. Vol.: | 10 µL splitless |
Oven | |
Oven Temp.: | 60 °C (hold 2 min) to 100 °C at 20 °C/min to 240 °C at 5 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 45 cm/sec |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 240 °C |
Source Temp.: | 230 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 70 eV |
Ionization Mode: | EI |
Scan Range: | 30-400 amu |
Notes | Inj: 10 µL/min Std. conc.: neat Gerstel CIS Injector: 35°C (hold 2 min.), to 300°C @ 10°C/sec. (hold 5 min.) Helium vent flow: 600mL/min with 1.8 min. vent end time |
Acknowledgement | Chromatogram courtesy of Kevin MacNamara, Ph.D., Irish Distilleries, Ltd. |