Peaks | |
---|---|
1. | Methanol |
2. | Ethanol |
3. | Isopropanol |
4. | tert-Butanol |
5. | n-Propanol |
6. | Isobutanol |
7. | Methyl Cellosolve® |
8. | n-Butanol |
Peaks | |
---|---|
9. | Cellosolve® |
10. | Ethylene glycol |
11. | Butyl Cellosolve® |
12. | Limonene |
13. | Dipropylene glycol monomethyl ether (DPGMME) |
14. | Diethylene glycol ethyl ether (DEGEE) |
15. | Phenyl Cellosolve® |
16. | Tripropylene glycol monomethyl ether (TPGMME) |
Column | Rtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916) |
---|---|
Standard/Sample | 1 % each solvent |
Diluent: | methanol |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 100:1) |
Liner: | Splitless (4mm ID) (cat.# 20814) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 60 °C (hold 1 min) to 138 °C at 4 °C/min to 145 °C at 2 °C/min to 210 °C at 15 °C/min to 230 °C at 5 °C/min |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.5 mL/min |
Detector | FID @ 250 °C |
---|