Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Butyl Cellosolve® |
2. | Limonene |
Column | Rtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916) |
---|---|
Standard/Sample | Approx. 5% cleaner in methanol |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Splitless (4mm ID) (cat.# 20814) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 60 °C (hold 1 min) to 138 °C at 4 °C/min to 145 °C at 2 °C/min to 210 °C at 15 °C/min to 230 °C at 5 °C/min |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.5 mL/min |
Detector | FID @ 250 °C |
---|
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.