Peaks | Conc. (wt.%) | |
---|---|---|
1. | Ethyl butyrate | 35.7 |
2. | Limonene | 20.0 |
3. | Eucalyptol | 0.5 |
4. | Geraniol | 0.6 |
5. | Benzoic acid | 1.0 |
6. | Cinnamic aldehyde | 0.5 |
Peaks | Conc. (wt.%) | |
---|---|---|
7. | Hydroxycitronellal | 5.0 |
8. | Thymol | 0.3 |
9. | Cinnamyl alcohol | 0.3 |
10. | Cinnamyl acetate | 0.3 |
11. | Vanillin | 0.1 |
12. | Benzyl salicylate | 35.7 |
Column | Rtx-1701, 60 m, 0.25 mm ID, 0.25 µm (cat.# 12026) |
---|---|
Standard/Sample | Fragrance materials test mix (cat.# 31807) |
Conc.: | 5% FMA mix in acetone |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 40:1) |
Liner: | Splitless (4 mm ID) (cat.# 20814) |
Inj. Temp.: | 285 °C |
Oven | |
Oven Temp.: | 50 °C to 270 °C at 3 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 0.6 mL/min |
Detector | FID @ 300 °C |
---|