Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | α-Pinene |
2. | Sabinene |
3. | β-Pinene |
4. | β-Myrcene |
5. | p-Cymene |
6. | D-Limonene |
7. | γ-Terpinene |
8. | Linalol |
9. | Linalyl acetate (bergamol) |
Column | Rtx-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12723) |
---|---|
Standard/Sample | bergamot oil |
Diluent: | acetone |
Conc.: | 5 % |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Inj. Temp.: | 280 °C |
Oven | |
Oven Temp.: | 40 °C (hold 0.5 min) to 300 °C at 5 °C/min |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 36 cm/sec, 6.77 psi, 46.7 kPa @ 40 °C |
Detector | Shimadzu QP 2010 MS |
---|---|
Transfer Line Temp.: | 280 °C |
Ionization Mode: | EI |
Scan Range: | 40-500 amu |
Instrument | Shimadzu 2010 GC & QP2010+ MS |
Acknowledgement | Chromatograms courtesy of Dr. Novalina Lingga, Shimadzu Asia Pacific, Singapore |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.