Peaks | |
---|---|
1. | β-Patchoulene |
2. | β-Elemene |
3. | Caryophyllene |
4. | α-Guaiene |
5. | Seychellene |
6. | α-Patchoulene |
7. | Guaiene |
8. | δ-Guaiene |
9. | Selinene |
10. | Patchouli alcohol |
Column | Rtx-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12723) |
---|---|
Standard/Sample | patchouli oil |
Diluent: | acetone |
Conc.: | 5 % |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Inj. Temp.: | 280 °C |
Oven | |
Oven Temp.: | 40 °C (hold 0.5 min) to 300 °C at 5 °C/min |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 36 cm/sec, 6.77 psi, 46.7 kPa @ 40 °C |
Detector | Shimadzu QP 2010 MS |
---|---|
Transfer Line Temp.: | 280 °C |
Ionization Mode: | EI |
Scan Range: | 40-500 amu |
Instrument | Shimadzu 2010 GC & QP2010+ MS |
Acknowledgement | Chromatograms courtesy of Dr. Novalina Lingga, Shimadzu Asia Pacific, Singapore |