Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | 5-α-Cholestane (IS) |
2. | Cholesterol |
Column | Rxi -5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420) |
---|---|
Standard/Sample | 1,000 μg/mL cholesterol in DMF |
1,000 μg/mL 5-α-cholestane in hexane | |
25 ng cholesterol, 150 ng 5-α-cholestane on column | |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 20:1) |
Liner: | Splitless taper (4 mm) w/wool (cat.# 22405) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 200 °C (hold 1 min) to 330 °C at 20 °C/min (hold 7.5 min) |
Carrier Gas | He, constant pressure (9.7 psi, 66.9 kPa) |
Linear Velocity: | 24 cm/sec @ 200 °C |
Detector | FID @ 340 °C |
---|
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.