Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Hexadecanoic acid |
Column | Rxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602) |
---|---|
Standard/Sample | Sweet potato spiked with pesticide mix and extracted with acetonitrile and Q-sep QuEChERS EN Method 15662 extraction salts |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1 min) |
Liner: | Single taper (4 mm) w/deact. wool (cat.# 22405) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 72.5 °C (hold 1 min) to 350 °C at 20 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
---|---|
Mode: | |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | TOF |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Instrument | Agilent/HP6890 GC |
Sample Preparation | A. Extract (without cleanup step) acidified with formic acid to pH 5. B. Extract with cleanup using Q-sep QuEChERS dSPE cleanup tube (cat.# 26124) acidified with formic acid to pH 5. |
Notes | m/z 60, 73, 87, 129, 256 plotted |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.