Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

QuEChERS Extract of Cannabis on Rxi-17Sil MS and Rxi-5ms by GCxGC-TOFMS

GC_FF1207
ColumnRxi-17Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123)
Rxi-5ms 0.9 m, 0.18 mm ID, 0.18 µm,
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:Rxi-17Sil MS: 40 °C (hold 1 min) to 320 °C at 5 °C/min (hold 3 min)
Rxi-5ms: 45 °C (hold 1 min) to 325 °C at 5 °C/min (hold 3 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:2 sec
Hot Pulse Time:0.6 sec
Cool Time between Stages:0.4 sec
DetectorTOFMS
Transfer Line Temp.:310 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:4.5 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-500 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
Sample Preparation Sample: Up to 2 g of finely ground cannabis in a 50 mL polypropylene centrifuge tube.

Extraction: Add 10 mL acetonitrile and 10 mL of organic-free water, then shake thoroughly. Soak 1 hour, then aggressively vortex 30 minutes. Add Q-sep packet (cat.# 26235), shake 1 minute, then centrifuge 5 minutes at 3000 g with Q-sep 3000 centrifuge (cat.# 26230).
NotesSee GC_FF1208 and GC_FF1209 to zoom in on terpenes and cannabinoids regions.