Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | Structural Nomenclature | |
---|---|---|---|
1. | Methyl myristate | 8.960 | C14:0 |
2. | Methyl palmitate | 11.753 | C16:0 |
3. | Methyl stearate | 15.545 | C18:0 |
4. | Methyl oleate | 16.716 | C18:1 cis-9 |
5. | Methyl vaccenate | 16.852 | C18:1 cis-11 |
6. | Methyl linoleate | 18.507 | C18:2 cis-9,12 |
7. | Methyl linolenate | 20.745 | C18:3 cis-9,12,15 |
Column | Rtx-2330, 105 m, 0.25 mm ID, 0.20 µm (cat.# 10729) |
---|---|
Standard/Sample | Butter flavored popcorn |
Diluent: | Hexane |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4 mm ID straight inlet liner w/wool (cat.# 23300) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 160 °C to 250 °C at 2 °C/min (hold 5 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2 mL/min |
Detector | FID @ 250 °C |
---|---|
Instrument | Agilent 7890A GC |
Sample Preparation | The sample was prepared using sodium methoxide. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.