Peaks | tR (min) | |
---|---|---|
1. | Acetaldehyde | 4.370 |
2. | Methanol | 4.474 |
3. | Ethanol | 5.395 |
4. | n-Propanol | 6.836 |
5. | Ethyl acetate | 7.403 |
6. | Isobutanol | 8.132 |
7. | Acetal | 9.458 |
8. | 3-Methyl-1-butanol (isoamyl alcohol) | 10.308 |
9. | 2-Methyl-1-butanol (active amyl alcohol) | 10.379 |
Column | Rxi-1301Sil MS, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16097) |
---|---|
Standard/Sample | Rum |
Conc.: | Neat |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 20:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 300 °C at 10 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Tune Type: | BFB | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | Water coelutes with acetaldehyde and methanol, which results in distortion of the baseline. |