Peaks | tR (min) | |
---|---|---|
1. | Tripalmitin (PPP) | 14.252 |
2. | 1,2-Palmitin-3-stearin (PPS) | 15.958 |
3. | 1,3-Palmitin-2-olein (POP) | 16.261 |
4. | 1,2-Palmitin-3-linolein (PPL) | 16.637 |
5. | 1-Palmitin-2-olein-3-stearin (POS) | 18.319 |
6. | 1,2-Olein-3-palmitin (POO) | 18.717 |
7. | 1-Palmitin-2-linolein-3-olein (PLO) | 19.159 |
8. | 1,2-Linolein-3-palmitin (PLL) | 19.617 |
9. | 1,2-Stearin-3-olein (SOS) | 20.752 |
10. | Triolein (OOO) | 21.496 |
Column | Rxi-65TG, 30 m, 0.25 mm ID, 0.1 µm (cat.# 17108) |
---|---|
Standard/Sample | Palm oil |
Conc.: | ~50 mg of palm oil was diluted to 10 mL with isooctane |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 35:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 360 °C |
Split Vent Flow Rate: | 35 mL/min |
Oven | |
Oven Temp.: | 200 °C (hold 1 min) to 350 °C at 20 °C/min to 365 °C at 1 °C/min (hold 6 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1 mL/min |
Detector | FID @ 365 °C |
---|---|
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 370 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890B GC |