Peaks | tR (min) | |
---|---|---|
1. | α-Pinene | 1.253 |
2. | 3-Methylcyclohexanone | 1.304 |
3. | β-Phellandrene | 1.367 |
4. | β-Pinene | 1.399 |
5. | p-Cymene | 1.551 |
6. | D-Limonene | 1.570 |
7. | Eucalyptol | 1.599 |
8. | Methone 1 | 2.231 |
Peaks | tR (min) | |
---|---|---|
9. | Methone 2 | 2.274 |
10. | Neoisomenthol | 2.295 |
11. | Menthol | 2.364 |
12. | Isomenthol | 2.417 |
13. | α-Terpineol | 2.440 |
14. | Pulegone | 2.706 |
15. | Menthyl acetate | 3.006 |
16. | Caryophyllene | 3.953 |
Column | Rxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602) |
---|---|
Standard/Sample | Peppermint oil |
Diluent: | Acetone |
Conc.: | 5% |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.01 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |