Peaks | tR (min) | |
---|---|---|
1. | α-Pinene | 0.514 |
2. | β-Myrcene | 0.560 |
3. | Octanal | 0.577 |
4. | α-Phellandrene | 0.597 |
5. | D-Limonene | 0.639 |
6. | γ-Terpinene | 0.671 |
7. | Linalool | 0.728 |
8. | Decanal | 0.942 |
9. | Z-Citral | 1.024 |
10. | Carvone | 1.051 |
Peaks | tR (min) | |
---|---|---|
11. | E-Citral | 1.089 |
12. | Geranyl acetate | 1.354 |
13. | Copaene | 1.391 |
14. | β-Cubebene | 1.417 |
15. | Caryophyllene | 1.512 |
16. | α-Caryophyllene | 1.604 |
17. | Germacrene D | 1.667 |
18. | Bicyclogermacrene | 1.704 |
19. | δ-Cadinene | 1.751 |
20. | Nootkatone | 2.468 |
Column | Rxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815) |
---|---|
Standard/Sample | Grapefruit oil |
Diluent: | Acetone |
Conc.: | 1% |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |