Peaks | tR (min) | |
---|---|---|
1. | α-Thujene | 0.502 |
2. | α-Pinene | 0.514 |
3. | Camphene | 0.537 |
4. | β-Pinene | 0.569 |
5. | α-Phellandrene | 0.596 |
6. | α-Terpinene | 0.611 |
7. | p-Cymene | 0.623 |
8. | Eucalyptol | 0.644 |
9. | γ-Terpinene | 0.672 |
10. | Terpinolene | 0.720 |
11. | Linalool | 0.728 |
Peaks | tR (min) | |
---|---|---|
12. | Camphor | 0.856 |
13. | Borneol | 0.902 |
14. | Terpinen-4-ol | 0.914 |
15. | α-Terpineol | 0.940 |
16. | Verbenone | 0.977 |
17. | Bornyl acetate | 1.148 |
18. | Eugenol | 1.310 |
19. | Copaene | 1.392 |
20. | Caryophyllene | 1.515 |
21. | α-Caryophyllene | 1.604 |
22. | γ-Muurolene | 1.643 |
23. | δ-Cadinene | 1.751 |
Column | Rxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815) |
---|---|
Standard/Sample | Rosemary oil |
Diluent: | Acetone |
Conc.: | 1% |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/Wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.01 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |