Peaks | tR (min) | |
---|---|---|
1. | α-Pinene | 0.514 |
2. | 3-Methylcyclohexanone | 0.533 |
3. | β-Phellandrene | 0.554 |
4. | β-Pinene | 0.568 |
5. | p-Cymene | 0.621 |
6. | D-Limonene | 0.630 |
7. | Eucalyptol | 0.639 |
Peaks | tR (min) | |
---|---|---|
8. | Methone 1 | 0.867 |
9. | Methone 2 | 0.887 |
10. | Menthol | 0.917 |
11. | Isomenthol | 0.936 |
12. | α-Terpineol | 0.943 |
13. | Pulegone | 1.044 |
14. | Menthyl acetate | 1.153 |
15. | Caryophyllene | 1.515 |
Column | Rxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815) |
---|---|
Standard/Sample | Peppermint oil |
Diluent: | Acetone |
Conc.: | 1% |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.01 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Inert | ||||||||
Source Temp.: | 230 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||
Notes | All peaks were identified using the NIST MS EI spectra library (2005). |