Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR 1 | tR 2 | |
---|---|---|---|
1. | (R,S)-Ethyl 2-methylbutyrate | 8.662 | 9.129 |
Column | Rt-βDEXse, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13106) |
---|---|
Standard/Sample | Ethyl-2-methylbutyrate |
Diluent: | Acetone |
Conc.: | 830 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305) |
Inj. Temp.: | 210 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 230 °C at 2 °C/min (hold 3 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 80 cm/sec @ 40 °C |
Detector | FID @ 230 °C |
---|---|
Constant Column + Constant Make-up: | 51.4 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |
Sample Preparation | Ethyl-2-methylbutyrate (neat) was dissolved in acetone to a final concentration of 830 ppm in 2 mL, screw-thread vials (cat.# 21143), and capped with short-cap, screw-vial closures (cat.# 24495). |
Notes | Ethyl-2-methylbutyrate enantiomeric resolution was 4.6. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.