Carvone on Rt-βDEXsa

GC_FF1369
PeakstR (min)Conc.
(mg/mL)
1.R(-)-Carvone36.6971
2.S(+)-Carvone40.2170.5
ColumnRt-βDEXsa, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13108)
Standard/SampleR(-)-Carvone
S(+)-Carvone
Diluent:Acetone
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:210 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 230 °C at 2 °C/min (hold 3 min)
Carrier GasH2, constant flow
Linear Velocity:80 cm/sec @ 40 °C
DetectorFID @ 230 °C
Constant Column + Constant Make-up:51.4 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Sample PreparationR(-)-Carvone and S(+)-carvone (neat) were dissolved in acetone to a final concentration of 1000 and 500 ppm (respectively) in 2 mL, screw-thread vials (cat.# 21143), and capped with short-cap, screw-vial closures (cat.# 24495).
NotesEnantiomeric resolution of carvone was 3.0.