Peaks | tR 1 | tR 2 | tR 3 | tR 4 | |
---|---|---|---|---|---|
1. | Rose oxides | 22.701 | 22.931 | 23.370 | 23.616 |
2. | (-) Menthone | 28.420 | - | - | - |
3. | Linalool (±) | 30.919 | 31.298 | - | - |
4. | β-Citronellol (±) | 37.090 | 37.245 | - | - |
5. | Geraniol | 39.494 | - | - | - |
Column | Rt-βDEXsa, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13108) |
---|---|
Standard/Sample | Geranium oil |
Diluent: | Acetone |
Conc.: | 5 % |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305) |
Inj. Temp.: | 210 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 230 °C at 2 °C/min (hold 3 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 80 cm/sec @ 40 °C |
Detector | FID @ 230 °C |
---|---|
Constant Column + Constant Make-up: | 51.4 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |
Sample Preparation | Geranium oil was dissolved in acetone to a final concentration of 5% in 2 mL, screw-thread vials (cat.# 21143), and capped with short-cap, screw-vial closures (cat.# 24495). |