Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Lactones on Rt-βDEXcst

GC_FF1378
PeakstR 1tR 2
1.(±)-γ-Nonalactone52.36553.038
2.(±)-γ-Undecalactone60.87961.399
3.(±)-δ-Decalactone63.46363.733
ColumnRt-βDEXcst, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13102)
Standard/Sample(±)-γ-Nonalactone
(±)-γ-Undecalactone
(±)-δ-Decalactone
Diluent:Acetone
Conc.:1000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:210 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 230 °C at 2 °C/min (hold 3 min)
Carrier GasH2, constant flow
Linear Velocity:80 cm/sec @ 40 °C
DetectorFID @ 230 °C
Constant Column + Constant Make-up:51.4 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Sample PreparationIndividual lactones (neat) were dissolved in acetone to a final concentration of 1000 ppm in 2 mL, screw-thread vials (cat.# 21143), and capped with short-cap, screw-vial closures (cat.# 24495).
NotesThe enantiomeric resolution of the individual lactones was (±)-γ-nonalactone 3.8, (±)-γ-undecalactone 3.1, and (±)-δ-decalactone 1.6.