Peaks | tR (min) | |
---|---|---|
1. | α-Pinene | 7.39 |
2. | Camphene | 7.71 |
3. | β-Myrcene | 7.98 |
4. | Sabinene | 8.02 |
5. | β-Pinene | 8.11 |
6. | α-Phellandrene | 8.4 |
7. | δ 3-Carene | 8.44 |
8. | α-Terpinene | 8.57 |
9. | Ocimene | 8.61 |
10. | Limonene | 8.71 |
11. | p-Cymene | 8.75 |
12. | β-Ocimene | 8.82 |
Peaks | tR (min) | |
---|---|---|
13. | Eucalyptol | 8.91 |
14. | γ-Terpinene | 9.06 |
15. | Terpinolene | 9.47 |
16. | Linalool | 9.87 |
17. | Fenchone | 10.06 |
18. | Isopulegol | 10.73 |
19. | dl-Menthol | 11.08 |
20. | Borneol | 11.19 |
21. | α-Terpineol | 11.29 |
22. | Dihydrocarveol | 11.40 |
23. | Citronellol | 11.51 |
24. | Geraniol | 11.82 |
Peaks | tR (min) | |
---|---|---|
25. | 2-Piperidinone | 11.88 |
26. | Citral 1 | 11.92 |
27. | Pulegone | 11.97 |
28. | Citral 2 | 12.24 |
29. | Citral 3 | 13.19 |
30. | Citral 4 | 13.43 |
31. | β-caryophyllene | 13.83 |
32. | α-Humulene | 14.21 |
33. | Nerolidol 1 | 14.78 |
34. | Nerolidol 2 | 15.08 |
35. | Caryophyllene oxide | 15.92 |
36. | α-Bisabolol | 16.43 |
Column | Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868) |
---|---|
Standard/Sample | Terpenes mix |
Diluent: | Isopropyl alcohol |
Conc.: | 200 ng/µL (0.02% wt/vol). The sample was prepared by placing 10 µL into the headspace vial. |
Injection | headspace-loop split (split ratio 10:1) |
Liner: | Premium 1.0 mm ID straight inlet liner (cat.# 23333.1) |
Headspace-Loop | |
Inj. Port Temp.: | 250 °C |
Instrument: | Tekmar HT-3 |
Inj. Time: | 1.0 min |
Transfer Line Temp.: | 160 °C |
Valve Oven Temp.: | 160 °C |
Needle Temp.: | 140 °C |
Sample Temp.: | 140 °C |
Sample Equil. Time: | 30.0 min |
Vial Pressure: | 20 psi |
Loop Pressure: | 15 psi |
Oven | |
Oven Temp.: | 60 °C (hold 0.10 min) to 300 °C at 12.50 °C/min (hold 3.0 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 33 cm/sec |
Detector | FID @ 320 °C |
---|---|
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | For qualitative purposes only. |