Peaks | tR (min) | |
---|---|---|
1. | Methamidophos | 8.20 |
2. | Dichlorvos | 8.61 |
3. | Mevinphos | 11.62 |
4. | Acephate | 11.66 |
5. | o-Phenylphenol | 12.93 |
6. | Omethoate | 14.07 |
7. | Dimethoate | 16.02 |
8. | gamma-BHC | 16.53 |
9. | Diazinon | 17.04 |
10. | Chlorothalonil | 17.28 |
11. | Vinclozolin | 18.21 |
12. | Carbaryl | 18.31 |
13. | Metalaxyl | 18.54 |
Peaks | tR (min) | |
---|---|---|
14. | Methiocarb | 18.88 |
15. | Pirimiphos methyl | 18.95 |
16. | Dichlofluanid | 19.09 |
17. | Malathion | 19.17 |
18. | Fenthion | 19.37 |
19. | Chlorpyrifos | 19.42 |
20. | Cyprodinil | 20.03 |
21. | Thiabendazole | 20.28 |
22. | Captan | 20.40 |
23. | Folpet | 20.56 |
24. | Imazalil | 21.49 |
25. | Myclobutanil | 21.83 |
26. | Endrin | 22.22 |
Peaks | tR (min) | |
---|---|---|
27. | Endosulfan sulfate | 23.42 |
28. | Fenhexamid | 23.47 |
29. | 4,4'-DDT | 23.50 |
30. | Propargite* | 23.92 |
31. | Iprodione | 24.41 |
32. | Dicofol | 24.83 |
33. | Bifenthrin | 24.72 |
34. | Fenpropathrin | 24.85 |
35. | Phosalone | 25.48 |
36. | Azinphos-methyl | 25.48 |
37. | cis-Permethrin | 26.83 |
38. | trans-Permethrin | 26.98 |
39. | Coumaphos | 27.10 |
40. | Deltamethrin | 29.95 |
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
---|---|
Standard/Sample | QuEChERS performance standards kit (cat.# 31152) |
Diluent: | Acetonitrile |
Conc.: | 1 ng/µL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 330 °C at 8 °C/min (hold 6.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 320 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 250 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 6.5 min | ||||||||
Tune Type: | PFTBA | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS |