Peaks | tR (min) | Conc. (ng/mL) | |
---|---|---|---|
1. | 3-MCPD-d5 PBA derivative | 4.143 | 100 |
2. | 3-MCPD PBA derivative | 4.169 | 200 |
3. | 2-MCPD-d5 PBA derivative | 4.388 | 100 |
4. | 2-MCPD PBA derivative | 4.418 | 100 |
Column | Rxi-17Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 14102) |
---|---|
Standard/Sample | See notes |
Diluent: | Isooctane |
Injection | |
Inj. Vol.: | 1 µL PTV split (split ratio 10:1) |
Liner: | Topaz 2.0 mm ID straight inlet liner w/wool (cat.# 23314) |
Inlet Temp. Program: | 120 °C to 165 °C at 300 °C/min (hold 10 min) to 320 °C at 300 °C/min (hold 8 min) |
Oven | |
Oven Temp.: | 120 °C (hold 0.5 min) to 200 °C at 18.5 °C/min to 330 °C at 35 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1 mL/min |
Detector | MS | |
---|---|---|
Mode: | SIM | |
SIM Program: | 147, 150, 196, 201 m/z, 50 ms dwell | |
Transfer Line Temp.: | 320 °C | |
Analyzer Type: | Quadrupole | |
Source Type: | Inert | |
Source Temp.: | 230 °C | |
Quad Temp.: | 150 °C | |
Ionization Mode: | EI | |
Instrument | Agilent 7890A GC & 5975C MSD | |
Sample Preparation | Standards were derivatized with 20 µL phenylboronic acid (saturated solution in diethyl ether), dried, and then reconstituted in 1 mL isooctane. Final concentrations are given in the peak table. |