Peaks | tR (min) | |
---|---|---|
1. | (E)-Arabinose-TMS-oxime | 21.180 |
2. | (Z)-Xylose-TMS-oxime | 21.916 |
3. | (E)-Ribose-TMS-oxime | 22.018 |
4. | (E)-Xylose-TMS-oxime | 22.075 |
5. | (Z)-Ribose-TMS-oxime | 22.186 |
6. | (E)-Rhamnose-TMS-oxime | 22.529 |
7. | (E)-Fucose-TMS-oxime | 22.662 |
8. | (Z)-Rhamnose-TMS-oxime | 22.735 |
Peaks | tR (min) | |
---|---|---|
9. | (Z)-Fucose-TMS-oxime | 23.072 |
10. | (E)-Fructose-TMS-oxime | 25.277 |
11. | (E)-Mannose-TMS-oxime | 25.643 |
12. | (E)-Galactose-TMS-oxime | 25.816 |
13. | (Z)-Mannose-TMS-oxime | 26.041 |
14. | (E)-Glucose-TMS-oxime | 26.066 |
15. | (Z)-Fructose-TMS-oxime | 26.329 |
16. | (Z)-Galactose-TMS-oxime | 26.562 |
17. | (Z)-Glucose-TMS-oxime | 26.653 |
Column | Rtx-225, 30 m, 0.25 mm ID, 0.25 µm (cat.# 14023) |
---|---|
Standard/Sample | Sugar TMS oximes |
Diluent: | Ethyl acetate |
Conc.: | 1-5 mg of starting material |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 400:1) |
Liner: | Topaz 4.0 mm ID Precision Inlet Liner w/ Wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 0.5 min) to 220 °C at 4.5 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Linear Velocity: | 43.4 cm/sec @ 40 °C |
Dead Time: | 1.14 min @ 40 °C |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Extractor | ||||||||
Source Temp.: | 280 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 978 eV | ||||||||
Solvent Delay Time: | 7 min | ||||||||
Tune Type: | PFTBA | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890B GC & 5977A MSD | ||||||||
Notes | Approximately 1 mg of sugar was dissolved in 100 μL of 40 mg/mL ethylhydroxylaminehydrochloride (EtOx) and heated at 70°C for 30 min. Samples were cooled to room temperature for approximately 5 min. 120 μL of N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA) was added to each sample and then heated at 70°C for 30 min. Samples were diluted 1:100 in ethyl acetate. |