Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Sugar alditol acetates on Rtx-1701

GC_FS0580
ColumnRtx-1701, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12023)
Standard/SampleAlditol Acetate Mixture-1 (Matreya, LLC, cat.# 1124)
Alditol Acetate Mixture-2 (Matreya, LLC, cat.# 1125)
Diluent:Chloroform
Conc.:50,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 400:1)
Liner:Topaz 4.0 mm ID Precision Inlet Liner w/ Wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 0.5 min) to 230 °C at 30 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:42.906 cm/sec @ 40 °C
Dead Time:1.165 min @ 40 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.550-7507.5
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Source Temp.:280 °C
Quad Temp.:150 °C
Electron Energy:978 eV
Solvent Delay Time:1.5 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD