Peaks | tR (min) | |
---|---|---|
1. | (E)-Fucose-TFA-oxime | 2.048 |
2. | (E)-Ribose-TFA-oxime | 2.132 |
3. | (E)-Arabinose-TFA-oxime | 2.176 |
4. | (Z)-Ribose-TFA-oxime | 2.320 |
5. | (Z)-Fucose-TFA-oxime | 2.347 |
6. | (E)-Xylose-TFA-oxime | 2.376 |
Peaks | tR (min) | |
---|---|---|
7. | (Z)-Arabinose-TFA-oxime | 2.473 |
8. | (Z)-Xylose-TFA-oxime | 2.555 |
9. | (E)-Mannose-TFA-oxime | 2.646 |
10. | (E)-Glucose-TFA-oxime | 2.719 |
11. | (Z)-Mannose-TFA-oxime | 2.830 |
12. | (Z)-Glucose-TFA-oxime | 2.894 |
Column | Rtx-1701, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12023) |
---|---|
Standard/Sample | Sugar TFA oximes |
Diluent: | Ethyl acetate |
Conc.: | 1-5 mg of starting material |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision Inlet Liner w/ Wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 150 °C (hold 1 min) to 280 °C at 28 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Linear Velocity: | 45.023 cm/sec @ 150 °C |
Dead Time: | 1.115 min @ 150 °C |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 280 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | Extractor | ||||||||
Source Temp.: | 280 °C | ||||||||
Quad Temp.: | 150 °C | ||||||||
Electron Energy: | 978 eV | ||||||||
Solvent Delay Time: | 1.5 min | ||||||||
Tune Type: | PFTBA | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890B GC & 5977A MSD | ||||||||
Notes | Approximately 1 of mg sugar was dissolved in 100 μL of 40 mg/mL ethylhydroxylaminehydrochloride (EtOx) and heated at 70°C for 30 min. Samples were cooled to room temperature for approximately 5 min. 120 μL of N-Methyl-bis(trifluoroacetamide) (MBTFA) was added to each sample and then heated at 70°C for 30 min. Samples were diluted 1:100 in ethyl acetate. |