Peaks | tR (min) | |
---|---|---|
1. | Diazinon | 13.97 |
2. | Isazophos | 14.27 |
3. | Chlorpyrifos methyl | 15.07 |
4. | Fenitrothion | 15.72 |
5. | Pirimiphos methyl | 15.76 |
6. | Chlorpyrifos | 16.21 |
7. | Pirimiphos ethyl | 16.71 |
8. | Quinalphos | 17.21 |
Peaks | tR (min) | |
---|---|---|
9. | Pyridaphenthion | 20.99 |
10. | Phosmet | 21.06 |
11. | EPN | 21.14 |
12. | Phosalone | 21.89 |
13. | Azinphos methyl | 21.90 |
14. | Pyrazophos | 22.58 |
15. | Azinphos ethyl | 22.63 |
16. | Pyraclofos | 22.73 |
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
---|---|
Standard/Sample | GC multiresidue pesticide standard #1-OPP (cat.# 32563) |
Diluent: | Toluene |
Conc.: | 100 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 290 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 325 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 5 min | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS | ||||||||
Notes | Reconstructed chromatogram from select ions. |