Peaks | tR (min) | |
---|---|---|
1. | Etridiazole | 9.31 |
2. | Atrazine | 13.29 |
3. | Terbuthylazine | 13.64 |
4. | Pyrimethanil | 13.85 |
5. | Terbacil | 14.11 |
6. | Vinclozolin | 15.06 |
7. | Triadimefon | 16.28 |
8. | MGK 264 1* | 16.56 |
9. | Cyprodinil | 16.77 |
10. | MGK 264 2* | 16.82 |
11. | Penconazole | 16.97 |
Peaks | tR (min) | |
---|---|---|
12. | Captan | 17.13 |
13. | Fipronil | 17.17 |
14. | Triadimenol | 17.21 |
15. | Folpet | 17.28 |
16. | Procymidone | 17.34 |
17. | Triflumizole | 17.41 |
18. | Paclobutrazol | 17.63 |
19. | Flutriafol | 17.90 |
20. | Tricyclazole (Beam) | 18.21 |
21. | Fludioxonil | 18.24 |
22. | Myclobutanil | 18.46 |
Peaks | tR (min) | |
---|---|---|
23. | Flusilazole | 18.53 |
24. | Bupirimate | 18.59 |
25. | Chlorfenapyr | 18.88 |
26. | Lenacil | 19.94 |
27. | Hexazinone (Velpar) | 20.24 |
28. | Tebuconazole | 20.31 |
29. | Propargite | 20.42 |
30. | Captafol | 20.48 |
31. | Iprodione | 20.88 |
32. | Pyriproxyfen | 21.97 |
33. | Fenarimol | 22.47 |
34. | Etofenprox | 24.53 |
35. | Fluridone (Sonar) | 24.92 |
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
---|---|
Standard/Sample | GC multiresidue pesticide standard #5-ONP (cat.# 32567) |
Diluent: | Toluene |
Conc.: | 100 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS | ||||||||
---|---|---|---|---|---|---|---|---|---|
Mode: | Scan | ||||||||
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 290 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 325 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 5 min | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Thermo Scientific TSQ 8000 Triple Quadrupole GC-MS | ||||||||
Notes | Propargite isomers are only slightly resolved with this method, so they are treated as one peak, but individual retention times are 20.40 and 20.42 min. Reconstructed chromatogram from select ions. |