Peaks | |
---|---|
1. | Ethyl acetate |
2. | n-Propyl acetate |
3. | MEK |
4. | Isopropyl alcohol |
5. | Ethanol |
6. | Isopropyl acetate |
7. | 2-Pentanone |
8. | MIBK |
9. | 1-Propanol |
Peaks | |
---|---|
10. | Butyl acetate |
11. | 1-Methoxy-2-propanol |
12. | p-Xylene |
13. | m-Xylene |
14. | Propylene glycol ethyl ether |
15. | o-Xylene |
16. | 1-Methoxy-2-propyl acetate |
17. | Cyclohexanone |
18. | Ethylene glycol monoethyl ether acetate |
19. | Diacetone alcohol |
Column | Stabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673) |
---|---|
Standard/Sample | Custom solvent standard |
Diluent: | Cyclohexane |
Conc.: | 400 ppm |
Injection | |
Inj. Vol.: | 0.5 µL split (split ratio 200:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 40 °C (hold 3 min) to 130 °C at 8 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS |
---|---|
Mode: | Scan |
Transfer Line Temp.: | 200 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 200 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 70 eV |
Solvent Delay Time: | 2.5 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Instrument | Agilent 7890A GC & 5975C MSD |
Notes | Scan Program: 30-150 amu |