Peaks | |
---|---|
1. | n-Propane |
2. | Isobutane |
3. | n-Butane |
4. | Methanol |
5. | n-Pentane |
6. | Ethanol |
7. | Diethyl ether |
8. | Acetone |
9. | 2-Propanol |
10. | Acetonitrile |
11. | Methylene chloride |
12. | n-Hexane |
Peaks | |
---|---|
13. | Ethyl acetate |
14. | Chloroform |
15. | Cyclohexane |
16. | Benzene |
17. | 1,2-Dichlorethane |
18. | n-Heptane |
19. | Trichloroethene |
20. | Toluene |
21. | m-Xylene |
22. | p-Xylene |
23. | o-Xylene |
24. | N,N-Dimethylacetamide (solvent) |
Column | Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868) |
---|---|
Standard/Sample | Residual solvents #1 (cat.# 34105) |
n-Propane, isobutane, n-butane (Emerald Scientific) | |
Diluent: | N,N-Dimethylacetamide |
Conc.: | 10 µg/mL |
Injection | Direct |
Liner: | Topaz 1.8 mm ID straight/SPME inlet liner (cat.# 23280) |
Inj. Temp.: | 280 °C |
Oven | |
Oven Temp.: | 30 °C (hold 6 min) to 85 °C at 15 °C/min (hold 2 min) to 250 °C at 35 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 2 mL/min |
Detector | FID @ 320 °C |
---|---|
Make-up Gas Flow Rate: | 40 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 45 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent 7890B GC |
Sample Preparation | A 6 mL sample for residual solvents analysis based on USP <467> was prepared for headspace solid phase microextraction (HS-SPME) as follows: 1. 3 g sodium chloride was weighed into a 20 mL amber headspace vial (cat. # 23086) with screw top cap (cat. # 23090). 2. 6 mL of deionized water was added to the vial. 3. The sample was fortified at 10 µg/mL with the residual solvents, n-propane, isobutane, and n-butane standards. 4. The vial was capped and vortexed at 3000 rpm for 10 seconds, inverted, then vortexed again for 10 seconds at 3000 rpm. |
Notes | CTC RTC Parameters HS-SPME Arrow Tool: SPME Arrow and SPME fiber SPME Arrow: Restek PAL SPME Arrow, DVB/C-WR/PDMS, 1.1 mm, 120 μm (cat.# 27875) SPME Fiber: Restek SPME fiber, DVB/C-WR/PDMS (cat.# 27873.1) Agitator Speed: 250 rpm Agitator Temp: 30 °C Incubation Time: 120 sec Heatex Stirrer Speed: 1000 rpm Heatex Stirrer Temp.: 30 °C Vial Penetration Depth: 35 mm Extraction Time: 120 sec Injector Penetration Depth: 50 mm Desorption Time: 10 sec Pre-Conditioning: True Post-Conditioning: False Conditioning Time: 60 sec Conditioning Temp.: 250 °C |