Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Terpenes MegaMix #1 on Rxi-1301Sil MS

GC_GN1211
PeakstR (min)
1.alpha-Pinene7.39
2.Camphene8.00
3.(-)-beta-Pinene9.07
4.Sabinene9.07
5.Beta-myrcene9.89
6.3-Carene10.30
7.R-(-)-alpha-Phellandrene10.30
8.alpha-Terpinene10.84
9.D-Limonene11.26
10.m-Isopropyltoluene11.32
11.p-Cymene11.51
12.Eucalyptol11.64
13.cis-beta-Ocimene11.86
14.o-Isopropyltoluene12.03
15.trans-beta-Ocimene12.37
PeakstR (min)
16.gamma-Terpinene12.56
17.Terpinolene13.74
18.Sabinene hydrate14.57
19.Linalool16.14
20.(1R)-endo-(+)-Fenchyl alcohol17.09
21.(-)-Isopulegol17.81
22.Isoborneol18.81
23.Terpinen-4-ol19.08
24.(+)-Borneol19.39
25.L(-)-Borneol19.39
26.Menthol19.52
27.alpha-Terpineol20.23
28.Nerol21.79
29.Citronellol21.99
30.Geraniol22.70
PeakstR (min)
31.Thymol24.87
32.Carvacrol25.08
33.(-)-alpha-Cedrene25.47
34.Beta-caryophyllene25.75
35.alpha-Humulene26.46
36.trans-beta-Farnesene26.46
37.Valencene27.12
38.cis-Nerolidol28.34
39.trans-Nerolidol28.81
40.(-)-Guaiol29.41
41.Cedrol29.53
42.(-)-alpha-Bisabolol30.66
43.Farnesol 130.92
44.Farnesol 231.25
45.Phytane (2,6,10,14-Tetramethylhexadecane)31.78
46.Squalene37.73
All compounds at 10 µg/mL.
ColumnRxi-1301Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16094)
Standard/SampleTerpenes MegaMix #1 (cat.# 34142)
Diluent:Isopropanol
Conc.:10 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 25:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:280 °C
Split Vent Flow Rate:45 mL/min
Oven
Oven Temp.:40 °C (hold 0.5 min) to 100 °C at 3 °C/min to 200 °C at 8 °C/min to 300 °C at 25 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:1.8 mL/min
Detector5977B
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
16.535-3502.3
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:HES
Source Temp.:325 °C
Quad Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:6.5 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977B MSD