Peaks | tR (min) | |
---|---|---|
1. | (1R)-(-)-Fenchone | 15.44 |
2. | (-)-alpha-Thujone | 16.45 |
3. | Camphor | 18.22 |
4. | l-Menthone | 18.61 |
5. | D-isomenthone | 19.11 |
6. | Safranal | 20.58 |
7. | Octyl acetate | 20.71 |
8. | (-)-Verbenone | 21.81 |
9. | d-Valerolactam | 21.92 |
Peaks | tR (min) | |
---|---|---|
10. | (R)-(+)-Pulegone | 22.07 |
11. | (S)-(+)-Carvone | 22.43 |
12. | cis-Citral | 22.58 |
13. | Piperitone | 22.97 |
14. | Isobornyl acetate | 23.18 |
15. | trans-Citral | 23.47 |
16. | Geranyl acetate | 25.47 |
17. | (-)-Caryophyllene oxide | 29.34 |
18. | (+)-Nootkatone | 32.97 |
Column | Rxi-1301Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16094) |
---|---|
Standard/Sample | |
Terpenes MegaMix #2 (cat.# 34143) | |
Diluent: | Isopropanol |
Conc.: | 10 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 25:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305) |
Inj. Temp.: | 280 °C |
Split Vent Flow Rate: | 45 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.5 min) to 100 °C at 3 °C/min to 200 °C at 8 °C/min to 300 °C at 25 °C/min (hold 3 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.8 mL/min |
Detector | 5977B | ||||||||
---|---|---|---|---|---|---|---|---|---|
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Type: | HES | ||||||||
Source Temp.: | 325 °C | ||||||||
Quad Temp.: | 200 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 6.5 min | ||||||||
Tune Type: | BFB | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Agilent 7890B GC & 5977B MSD |